RKI Hygiene Room per KRINKO

Meet RKI requirements – free of charge

Use the free ClavioSoft to run an audit check of your practice in just a few minutes – the software automatically generates any missing documentation. This saves you up to 90 % of the work.

Clavio has an idea
 

Setting Up a Hygiene Room per KRINKO

 

The reprocessing room is the centrepiece of instrument reprocessing. KRINKO requires a spatial-functional separation of reprocessing into the areas/zones unclean, clean and storage, with non-crossing material and personnel flow. The term black-and-white principle, commonly used in practice, does not appear in the KRINKO text itself.

1Unclean and Clean Area

Unclean area – for receiving, cleaning, disinfection and pre-treatment of contaminated instruments:

· Stations for manual cleaning/disinfection · Washer-disinfectors (WD) · Ultrasonic cleaners.

Clean area – for visual inspection, care, packaging and sterilisation:

· Autoclaves · Workstations for visual inspection and packaging. Only clean, disinfected instruments may be processed here.

Storage – a dedicated zone for recontamination-protected storage of packaged sterile goods:

· Cabinets for sterile goods. According to KRINKO (Annex 5), storage must be maintained as a separate zone and is not to be counted as part of the clean area.

Sources: KRINKO/BfArM 2012, Sect. 1.2/2.2 and Annex 5 (Requirements for Reprocessing Units, AEMP).

2Spatial or Temporal Separation

The required degree of separation depends on the category of the reprocessing unit: Category A (semi-critical A / critical A) permits a zone separation of unclean – clean – storage, where temporal separation is possible; Category B (semi-critical B / critical B) requires a spatial area separation into unclean – clean – storage; Category C requires dedicated rooms for each. Temporal separation alone therefore only suffices for Category A. The decisive factor is always that contaminated and sterile instruments must not come into contact.

Practical note: Even a single room can be KRINKO-compliant if it is divided into zones and workflows are separated temporally/organisationally. What matters are clear, non-crossing paths from unclean to clean.

3Workflows & Hygiene Requirements

· Non-crossing paths: contaminated and sterile instruments must not cross; areas must be clearly marked.
· Surfaces: smooth, joint-sealed and designed so that they can be easily cleaned and wipe-disinfected and are resistant to the (listed) surface disinfectants used. The KRINKO recommendation does not specify particular materials; it refers to the recommendation on surface cleaning and disinfection.
· Hygiene plan: regular cleaning/disinfection of all work areas according to a defined plan.

Clean area, safe sterilisation: the right autoclave is the centrepiece of your reprocessing room. Find the device that suits your practice.
Find the right Class B autoclave

Important Additional Points per RKI & Standards

These points are frequently overlooked in practice but are essential for correct, legally compliant implementation – and are precisely the aspects checked during a practice inspection:

  • Recontamination-protected storage and sterile-goods labelling: Packaged sterile goods must be stored in the storage zone in a dust-protected, clean, dry environment at room temperature to prevent recontamination (e.g. in closed cabinets/drawers). The packaging must be labelled (including sterilisation date, batch identification and, where applicable, expiry date and sterile-goods shelf life if shorter than the expiry date). The storage period depends on packaging, seal integrity and storage conditions.Source: KRINKO/BfArM 2012, Sect. 2.2.6 (Labelling) and Sect. 3 (Transport and Storage)
  • Qualified personnel: Reprocessing may only be carried out by personnel with documented specialist knowledge. This includes instrument science, knowledge of hygiene/microbiology including transmission routes, as well as risk assessment and classification of medical devices and the procedures used. Specialist knowledge requirements are described in Annex 6 of the KRINKO/BfArM recommendation (in conjunction with MPBetreibV).Source: KRINKO/BfArM 2012, Annex 6 (Specialist Knowledge of Personnel); MPBetreibV
  • Validated processes and batch documentation: Validated processes must be used throughout reprocessing; the recorded process parameters and the release decision must be documented with reference to the releasing person and the batch (batch documentation). Records must be retained for at least 5 years and may also be kept on data carriers; digital documentation software facilitates traceable recording and release – for example the free ClavioSoft, which lets you complete batch and release documentation in just a few clicks and store it in an audit-proof manner.Source: KRINKO/BfArM 2012, Sect. 2.2.8 (Batch Documentation); MPBetreibV
  • Non-crossing material flow and drying before transfer to the clean area: Material flow must be consistently unidirectional (unclean -> clean -> storage) without return paths. Reprocessing includes cleaning, optional intermediate rinsing, disinfection, rinsing and drying; complete drying is a prerequisite for packaging and sterilisation. The reprocessing instructions provided by the manufacturer (per DIN EN ISO 17664) must be followed. Contaminated instruments must be transported in safely enclosed (closed transport containers) to the reprocessing location.Source: KRINKO/BfArM 2012, Sect. 1.2 and Annex 5 (Spatial-Functional Separation) and Sect. 2.2.1/2.2.2 (Preparation, Cleaning/Drying)
  • When a separate room is not mandatory (Category A): A dedicated reprocessing room is not mandatory in every case: for simple Category A reprocessing, the process may take place in a single room divided into the zones unclean - clean - storage, provided workflows (including temporal ones) are properly separated and paths are non-crossing. For Categories B and C, this is not sufficient.Source: KRINKO/BfArM 2012, Annex 5 (Category A: Zone Separation, Temporal Separation Possible; Footnote 5: A Room May Be Divided into Areas and Zones)

Required Equipment, Consumables & Accessories

Checklist for this step – items you should have on hand (shopping list):

Equipment

  • Washer-disinfector (WD / thermal disinfector) for validated automated cleaning and disinfection (unclean area) Order here in the shop
  • Ultrasonic cleaner for pre-cleaning hard-to-reach and hinged instruments (unclean area) Order here in the shop
  • Class B autoclave (steam tabletop steriliser with fractionated vacuum) for sterilisation (clean area) Order here in the shop
  • Sealing device for standards-compliant packaging of sterile goods (clean area) Order here in the shop
  • Water treatment system for demineralised/distilled feed water for the autoclave Order here in the shop
  • Suitable test instrumentation for routine testing (e.g. process challenge device / PCD) Order here in the shop
  • Separate hand-wash station with soap dispenser, single-use towel dispenser and hand disinfectant dispenser (unclean area)
  • Lockable cabinets/drawers for recontamination-protected sterile-goods storage (storage zone)
  • Digital documentation software for batch and release documentation Order here in the shop Create digitally with ClavioSoft

Consumables

  • Sterilisation packaging (see-through pouches and rolls) for standards-compliant packaging Order here in the shop
  • Instrument cleaner and disinfectant (listed) for manual cleaning and WD process chemistry
  • Surface disinfectant with proven efficacy (listed) for work surfaces and surrounding areas
  • Hand disinfectant and liquid soap
  • Fluid-resistant disposable gloves and single-use protective gowns/aprons as PPE (unclean area)
  • Chemical indicators or process challenge devices for routine testing of sterilisation Order here in the shop
  • Instrument care lubricant for hinged instruments before sterilisation Order here in the shop
  • Demineralised/distilled water or filter cartridges for water treatment Order here in the shop
  • Labelling material/labels (sterilisation date, batch, shelf life/expiry date where applicable) Order here in the shop

Accessories

  • Closed, labelled transport containers for contaminated instruments (unclean area) Order here in the shop
  • Joint-sealed, wipe-disinfectable work surfaces, separated for unclean and clean areas
  • Hands-free waste container for contaminated waste
  • Instrument trays, perforated cassettes and loading baskets for WD and autoclave Order here in the shop
  • Visual inspection aid (magnifier/lighting) for inspecting cleaned instruments
  • Lint-free cloths or compressed air for residual moisture removal (drying)
  • Zone signage/labelling (unclean - clean - storage) and non-crossing routing
  • Written hygiene plan and standard operating / reprocessing procedures Create digitally with ClavioSoft

References

  1. KRINKO/BfArM 2012, Bundesgesundheitsbl. 55:1244–1310 – Sect. 1.2/2.2; Annex 5 (AEMP).
  2. MPBetreibV § 8.

How audit-ready is your practice?

Find out in just a few minutes how well prepared you are for the next inspection — category by category, with clear progress tracking and specific open items. Completely free with ClavioSoft.

ClavioSoft Audit-Ready Check: progress display and categories
Download ClavioSoft for free
Free forever For autoclaves from all manufacturers RKI-compliant

1 to 1 (from a total of 1)

Frequently Asked Questions

RKI Hygiene Room per KRINKO – key questions answered clearly.

What is a hygiene room per KRINKO?
The hygiene or reprocessing room is the area where medical devices are reprocessed. KRINKO requires a spatial-functional separation into the zones unclean, clean and storage with non-crossing material and personnel flow.
Which areas must a reprocessing room have per KRINKO?
Three zones: an unclean area for receiving, cleaning and disinfection, a clean area for visual inspection, care, packaging and sterilisation, and a dedicated storage zone for packaged sterile goods. The basis is KRINKO/BfArM 2012 (Annex 5).
What belongs in the unclean area of instrument reprocessing?
The unclean area houses the receiving, manual cleaning and disinfection of contaminated instruments, along with the associated equipment: washer-disinfectors (WD / thermal disinfector) and ultrasonic cleaners. Only instruments that have not yet been reprocessed are handled here.
What belongs in the clean area of the hygiene room?
The clean area houses visual inspection, care, packaging and sterilisation, i.e. the workstations for visual inspection and packaging as well as the autoclave. Only clean, disinfected instruments may be processed there.
What is the black-and-white principle in instrument reprocessing?
The black-and-white principle is the common practical term for separating the unclean (contaminated) and clean areas. The KRINKO text itself does not use this term; instead it refers to spatial-functional separation into unclean, clean and storage.
Does the autoclave have to be in a separate room?
A separate room is not mandatory; the autoclave is located in the clean area and may also be placed in the treatment room. If reprocessing takes place in the treatment room, this is only permitted when no patients are present. A window and fixed water connection are not required.
Is a single room sufficient for KRINKO-compliant reprocessing?
Yes, even a single room can be KRINKO-compliant if it is divided into the zones unclean, clean and storage and workflows are properly separated. What matters are clear, non-crossing paths so that contaminated and sterile instruments never come into contact.
What does non-crossing material flow mean in reprocessing?
Non-crossing material flow means that instruments are consistently moved in one direction: unclean to clean to storage, without return paths. Contaminated and sterile instruments must not cross; the areas must be clearly marked.
What spatial requirements apply to autoclave validation?
The KRINKO requirements apply, in particular the separation of clean and unclean areas. A drain hose is needed but may also lead into a container; a sink is not mandatory. A water supply connection is not required thanks to the integrated tank.
What distinguishes Category A, B and C reprocessing units?
The degree of separation depends on the category: Category A permits zone separation with possible temporal separation, Category B requires spatial area separation into unclean, clean and storage, and Category C requires dedicated rooms for each. The basis is KRINKO/BfArM 2012.
Must every sterilisation cycle be documented and released?
Yes. For validated processes, the process parameters of every batch must be recorded and the release documented with reference to the batch and the releasing person. Only after this batch release is the sterile goods considered fit for use. The basis is KRINKO/BfArM 2012 and MPBetreibV.
What does temporal instead of spatial separation mean in the reprocessing room?
For Category A, unclean and clean work steps may be performed sequentially on the same surfaces, with cleaning in between. From Category B onwards, permanent spatial area separation is required. Contaminated and sterile instruments must never come into contact.
Does sterile-goods storage belong to the clean area?
No. According to KRINKO (Annex 5), storage must be maintained as a separate zone and is not to be counted as part of the clean area. It serves the recontamination-protected storage of packaged sterile goods in closed cabinets or drawers.
How must packaged sterile goods be stored?
Packaged sterile goods must be stored in the storage zone in a dust-protected, clean, dry environment at room temperature to prevent recontamination, for example in closed cabinets or drawers. The basis is KRINKO/BfArM 2012, Section 3 (Transport and Storage).
What requirements apply to surfaces in the reprocessing room?
Surfaces must be smooth, joint-sealed and easy to clean and wipe-disinfect, and resistant to the listed surface disinfectants used. KRINKO does not specify particular materials; it refers to the recommendation on surface cleaning and disinfection.
Who must maintain and initially validate the autoclave?
An initial validation is required before commissioning, followed by regular revalidation and maintenance by qualified service personnel. These confirm that the process sterilises reproducibly. We carry out initial validation, maintenance and revalidation throughout Germany and in AT/CH.
Do I need a hygiene plan for the reprocessing room?
Yes. KRINKO requires regular cleaning and disinfection of all work areas according to a defined hygiene plan. The plan specifies which surfaces are to be cleaned, how often and with which agents.
Why must instruments be completely dried before entering the clean area?
Complete drying is a prerequisite for packaging and sterilisation. Residual moisture compromises the seal, the sterilisation result and shelf life. Only after cleaning, disinfection, rinsing and drying may the instrument pass into the clean area.
How must contaminated instruments be transported to the reprocessing room?
Contaminated instruments must be transported safely enclosed in closed transport containers to the reprocessing location. This prevents contamination spread and injury risks. The basis is KRINKO/BfArM 2012.
Who is permitted to reprocess instruments in a practice?
Reprocessing may only be carried out by personnel with documented specialist knowledge. This includes instrument science, knowledge of hygiene and microbiology including transmission routes, as well as risk assessment and classification of medical devices. This is described in Annex 6 of the KRINKO/BfArM recommendation in conjunction with MPBetreibV.
What specialist knowledge must reprocessing personnel demonstrate?
Required knowledge includes instrument science, hygiene and microbiology including transmission routes, as well as the ability to perform risk assessment and classification of medical devices and knowledge of the procedures used. The requirements are set out in Annex 6 of the KRINKO/BfArM recommendation.
Do I need training to operate the autoclave?
An introduction to the medical device is required; a trained, instructed and qualified person is sufficient. The introduction can be booked separately or is already included with the initial validation, installation and commissioning.
Must validated processes be used in reprocessing?
Yes. Validated processes must be used throughout reprocessing. The recorded process parameters and the release decision must be documented with reference to the releasing person and the batch. The basis is KRINKO/BfArM 2012 and MPBetreibV.
How long must batch documentation be retained?
Reprocessing records must be retained for at least 5 years and may also be kept on data carriers. Digital software such as the free ClavioSoft facilitates traceable, audit-proof batch and release documentation.
What must be labelled on sterile-goods packaging?
The packaging must include the sterilisation date, batch identification and, where applicable, expiry date and sterile-goods shelf life if shorter than the expiry date. This ensures traceability. The basis is KRINKO/BfArM 2012, Section 2.2.6.
What determines the storage period of packaged sterile goods?
The storage period depends on the packaging type, the integrity of the seals and the storage conditions. With dust-protected, dry storage in closed cabinets, longer shelf lives are possible. KRINKO does not specify a blanket time limit.
Do I need a water connection in the hygiene room?
No, a fixed water connection is not strictly necessary. The autoclaves are stand-alone devices with an integrated fresh and waste water tank and can be manually filled with distilled or demineralised water. A fixed water connection is optional and can be retrofitted.
Can the autoclave waste water drain into the sink?
Yes, the included hose can be mounted at the rear of the device and led into a container, a drain or the sink. A fixed waste-water connection is not required. Ensure that the hose end does not come into contact with the waste water.
What water does the autoclave need for steam sterilisation?
Demineralised or distilled feed water is required for steam generation; tap water is not suitable and will damage the device. A water treatment system provides the required quality. Feed-water requirements are specified in DIN EN 13060.
Is a standard power socket sufficient for the equipment in the reprocessing room?
Yes, a standard household socket is sufficient; three-phase power is not required. For the model without a fixed water connection, no additional connections beyond the power supply are needed.
Do you offer planning assistance for setting up an in-house sterilisation unit?
Yes, we are happy to advise you on planning your reprocessing and sterilisation setup, from equipment selection through installation to validation and maintenance. Contact us for individual consultation.
Which equipment belongs in a fully equipped hygiene room?
The unclean area requires an ultrasonic cleaner and optionally a thermal disinfector/WD; the clean area requires workstations for visual inspection and packaging with a sealing device and a Class B autoclave. The storage zone houses closed cabinets for sterile goods.
What is a sterile-goods packing workstation (SPA)?
SPA stands for sterile-goods packing workstation (Sterilgut-Pack-Arbeitsplatz). This is where reprocessed instruments are inspected, maintained, assembled and packaged after cleaning and disinfection. It is located in the clean area, before sterilisation.
Which autoclave belongs in the clean area of the hygiene room?
For critical instruments, the clean area requires a Class B steam steriliser per DIN EN 13060, the centrepiece of the reprocessing room. Use our configurator to find the right volume and features for your practice.
Do I need separate work surfaces for clean and unclean areas?
Yes, KRINKO requires spatial-functional separation. Receiving, cleaning and disinfection take place on unclean-area surfaces; visual inspection, packaging and sterilisation in the clean area. Both areas must be clearly marked and arranged in a non-crossing layout.
How much space does a KRINKO-compliant reprocessing room need?
KRINKO does not specify a minimum floor area. What matters is that the zones unclean, clean and storage can be accommodated with non-crossing material and personnel flow and sufficient work surface. Even a small, well-organised room can be compliant.
Does the hygiene room need a window or ventilation?
KRINKO does not require a window. What matters are smooth, easy-to-clean surfaces, a hygiene plan and proper zone separation. Neither a window nor a fixed water connection is required for equipment operation itself.
What does spatial-functional separation mean?
Spatial-functional separation means that the functions of receiving/cleaning, visual inspection/packaging/sterilisation and storage are assigned to separate zones and the material flow runs in one direction without crossing. Depending on the category, temporal separation may suffice or spatial separation is required.
Why must contaminated and sterile instruments never come into contact?
If contaminated and sterile instruments come into contact, there is a risk of recontamination of already reprocessed sterile goods. This is why KRINKO requires non-crossing flow from unclean to clean to storage without return paths.
How is personnel flow managed in the reprocessing room?
In addition to material flow, KRINKO requires non-crossing personnel flow. Paths from unclean to clean must be unambiguous, and when moving between areas, hand hygiene and glove changes where applicable must be observed to prevent contamination spread.
What steps are part of instrument reprocessing in the hygiene room?
Reprocessing includes cleaning, optional intermediate rinsing, disinfection, rinsing and drying, followed by visual inspection, care, packaging and steam sterilisation. The precise requirements are based on KRINKO/RKI and the manufacturer's reprocessing instructions for the instruments.
What lighting and working conditions does the visual inspection workstation need?
Good, glare-free illumination is important for visual inspection so that residual contamination and damage can be detected; magnifying lamps may help. The workstation is located in the clean area and requires sufficient space for care and packaging.
How much does it cost to set up a KRINKO-compliant hygiene room?
Costs depend on the equipment needed, such as an ultrasonic cleaner, optionally a thermal disinfector, a sealing device and a Class B autoclave. Use our configurator to put together a suitable complete package; construction measures for smooth surfaces and zone separation are additional.
Is a thermal disinfector mandatory in the hygiene room?
For critical-B instruments with lumens, automated cleaning in a WD / thermal disinfector is strongly preferred as it is standardised and validatable. For critical-A instruments, an ultrasonic cleaner may suffice. The risk classification and validated process are decisive.
What maintenance and performance verification records are required in the hygiene room?
In addition to batch documentation, records of the initial validation, regular revalidation and maintenance and performance verification of the autoclave must be retained. They demonstrate proper operation to regulatory authorities. The obligation arises from KRINKO/BfArM 2012 and MPBetreibV.
How often must the hygiene room be cleaned and disinfected?
All work areas must be cleaned and disinfected regularly according to the defined hygiene plan. Work surfaces are typically wipe-disinfected daily or upon contamination. Use only listed surface disinfectants.
How do I verify the sterilisation performance of a Class B autoclave?
Performance is verified through routine checks such as vacuum test, steam penetration test, and chemical and, where applicable, biological indicators, supplemented by periodic revalidation. This documents the effectiveness of every batch. The basis is DIN EN 13060 and the validation requirements.
Does the KRINKO hygiene room also apply to podiatry and foot care?
Yes, KRINKO requirements apply across all sectors where medical devices are reprocessed, including podiatry and foot care. Zone separation of unclean/clean/storage, non-crossing flow and a qualified person are required here as well.
Do the hygiene room requirements also apply to tattoo and piercing studios?
Yes. If instruments are reused, they are generally classified as critical and must be sterilised in packaging; accordingly, reprocessing with zone separation of unclean/clean/storage and qualified personnel is required. The exact classification follows KRINKO/RKI.
Do the room requirements differ in Austria and Switzerland?
The basic principle of zone separation into unclean/clean/storage is comparable in DE, AT and CH. However, country-specific regulations apply; in Austria, for example, the OENorm, and in Switzerland specific validation and hygiene requirements. Always check the applicable national requirements.