RKI Instrument Packaging

Meet RKI requirements – free of charge

Use the free ClavioSoft to run a complete audit check of your practice in just a few minutes – the software automatically generates any missing documentation. This saves you up to 90 % of the work.

ClavioSoft mascot Clavio: audit check and sterile packaging documentation per RKI
 

Sterile Packaging: How to Correctly Package Instruments

 

Packaging critical instruments preserves their sterility until the point of use. It forms the sterile barrier system (SBS), which keeps out dust, moisture and micro-organisms after sterilization. Packaging materials must meet the requirements of DIN EN ISO 11607-1 (or the DIN EN 868 series) and must be declared suitable by the manufacturer for the chosen sterilization method (e.g. steam at 134 degrees C). The sealing process must be validated in accordance with DIN EN ISO 11607-2.

1What is a sterile barrier system (SBS)?

An SBS is the sealed packaging that protects the sterile instrument from recontamination until the point of use (e.g. a transparent sterilization pouch or a container). It must form a microbial barrier, allow steam penetration, and permit aseptic opening. Requirements for materials and processes are governed by DIN EN ISO 11607-1 (materials/systems) and -2 (validation of the packaging process).

Sources: KRINKO/BfArM 2012, Section 2.2.4; DIN EN ISO 11607-1/-2; DGSV guideline on validation of packaging processes.

2Common packaging types

Packaging typeSuitability & requirements
Sterilization paper rolls (with a sealing device)Individually cuttable to size; ideal for larger instruments/sets. Wrinkle-free seal seam (≥ 6 mm wide), distance from seal seam to instrument ≥ 30 mm (3 cm); if the manufacturer specifies higher values, the manufacturer's specification applies.
Self-sealing pouchesConvenient (simply peel and seal), contents visible through transparent film; for smaller instruments. Wrinkle-free seam, overlap ≥ 1 cm, labelling with a solvent-free marker outside the seal seam.
Sterilization containersProvide excellent protection for sets/surgical instruments. Wrap instruments in lint-free cloth; replace disposable filters before each use, maintain and document reusable filters; keep steam ports clear; seal/lock/tamper-proof the container for storage.
Crepe/non-woven sheetsFlexible, ideal for irregularly shaped instruments and mixed sets; use protective caps to prevent puncture by sharp/pointed tips.

3The seal seam – and why it must be inspected

The seal seam is the most critical point of the packaging. It must be continuous, wrinkle-free and firm. A validated sealing process (temperature, contact pressure, speed) must be demonstrated in accordance with DIN EN ISO 11607-2. During routine operation, inspect the seam regularly:

Daily: Seal check / ink test, visual inspection of the seam (uniform contrast, sharp contours) and peel test (paper separates from the film with virtually no residue).
Periodically: Tensile strength testing of the seam.

Double sterile barrier system: For particularly sensitive instruments or long-term storage, a second outer wrap (double SBS / protective packaging) may be advisable. Do not overfill packages before sterilization – steam must be able to reach all surfaces.
Sources: DGSV guideline on validation of packaging processes (DIN EN ISO 11607-2); routine operation checklist for sealing devices; DIN 58953 (sterile supply management).
Reliable packaging starts with a properly sealing device and the right autoclave. Let us put together a matched setup for you.
Find the right Klasse B autoclave

Important additional points per RKI & standards

These points are frequently overlooked in practice but are essential for correct and legally compliant implementation:

  • Process/treatment indicator on the packaging: Use sterilization packaging with an integrated process indicator (treatment indicator) or apply one separately. The colour change confirms that the package has been through the sterilization process and distinguishes processed from unprocessed sterile goods. Visual inspection of the indicator is part of the batch-related checks and the release procedure.Source: KRINKO/BfArM 2012, Appendix 'Notes on the use of chemical indicators' (Class 1 process indicators) and Section 2.2.6 (process indicators for labelling)
  • Labelling in the course of sterilization and release: During packaging, non-critical preliminary labelling (e.g. contents) may be applied outside the seal seam and only on the paper side. The labelling-relevant information – sterilization date or batch identification, expiry date or sterile shelf life where applicable, and the release decision – is documented on the packaging in the course of sterilization and release, not in advance during packing. This batch- and release-related documentation (batch documentation) can be completed in just a few clicks with the free ClavioSoft – stored in a tamper-proof format.Source: KRINKO/BfArM 2012, Section 2.2.6 (labelling: sterilization date, batch identification, expiry date, sterile shelf life, release decision)
  • Mandatory inspection of the packaging immediately before use: Regardless of batch-related checks, the packaging must be inspected again immediately before each use (dryness, integrity, intact seal seams, valid labelling). If the packaging is damaged, moist or the labelling is invalid, the instrument must not be used but must be reprocessed and repackaged.Source: KRINKO/BfArM 2012, Section 2.2.4 ('Obligation to inspect immediately before use') and Annex 4 (batch-related checks – visual inspection of packaging: dryness, integrity, intact seal seams, complete labelling)
  • Sterile shelf life depends on packaging and storage conditions: Sterile goods must be stored dust-free, clean, dry and at room temperature. The permissible storage period depends on the quality of the packaging material, the tightness of the seal seams and the storage conditions; under favourable conditions, shelf lives of more than six months are possible. Define the shelf life within your facility and take into account the information provided by the manufacturers of both the medical device and the packaging material.Source: KRINKO/BfArM 2012, Section 3 (transport and storage): dust-free/clean/dry, storage period depends on packaging quality, seal tightness and storage conditions, shelf lives of more than six months feasible
  • Dryness of sterile goods before packaging: Only instruments that have been thoroughly cleaned, disinfected and dried may be packaged. Residual moisture under the film compromises the sterile barrier and shelf life and may cause corrosion. Before packaging, check cleanliness, dryness and functionality (e.g. place hinged instruments in the open position).Source: KRINKO/BfArM 2012, Sections 2.2.1/2.2.4 (inadequate drying as a source of error; visual check for dryness) and Annex 4 (place instruments with open hinges/joints)
  • Packaging process as a validated, SOP-governed part of the overall process: The packaging step is part of the reprocessing procedure that, under the MPBetreibV, must be carried out using suitable validated methods and by personnel with the required expertise, and must be defined in a standard operating procedure (SOP). Heat sealers must be validated and maintained in accordance with the manufacturer's instructions; the critical parameters (in particular temperature and contact pressure) must be set and documented. The SOP and ongoing documentation can be managed quickly and easily with the free ClavioSoft.Source: MPBetreibV (reprocessing with validated methods); KRINKO/BfArM 2012, Annex 4 (heat sealers: critical process parameters temperature and contact pressure, operating instructions must be available; prepare standard operating procedure); DIN EN ISO 11607-2

Required equipment, consumables & accessories

Checklist for this step – items you should have on hand (shopping list):

Equipment

  • Validatable heat sealer with adjustable temperature and contact pressure; for higher throughput, a continuous sealing device with automatic cutting function Order in our shop
  • Klasse B autoclave (steam sterilizer per DIN EN 13060) as the subsequent process step; the packaging material must be suitable for its procedure Order in our shop
  • Packing station in the clean zone with good lighting for visual inspection of the seal seam

Consumables

  • Self-sealing transparent sterilization pouches in various sizes for small individual instruments Order in our shop
  • Transparent sterilization rolls (paper/film) for custom cutting to size for larger instruments and sets Order in our shop
  • Crepe/non-woven sterilization sheets for irregularly shaped instruments and mixed sets
  • Container filters (replace disposable filters before each use or maintain reusable filters) for sterilization containers Order in our shop
  • Lint-free cloths for wrapping set instruments inside the container
  • Solvent-free, rub-resistant sterilization markers (label only on the paper side, outside the seam) Order in our shop
  • Packaging with integrated process indicator or separate process indicators Order in our shop
  • Protective caps or protective paper for sharp/pointed instrument tips (autoclavable) Order in our shop

Accessories

  • Seal-check / ink-test strips and peel-test material for daily seal seam inspection Order in our shop
  • Test equipment or service for periodic tensile strength testing of the seam (as part of maintenance/revalidation) Order in our shop
  • Validation and maintenance contract for the sealing device (initial validation, interval maintenance per manufacturer specifications) Order in our shop Request validation
  • Sterilization containers with lid/filter holder and locking/latching system for storage and transport Order in our shop
  • Sample tray list or packing schema template for documenting the contents and layout of each tray/set Create digitally with ClavioSoft
  • Standard operating procedure 'Packaging' as well as documentation/release software or forms Order in our shop Create digitally with ClavioSoft
  • Dust-proof, enclosed storage cabinet or drawer for packaged sterile goods

Templates & sources

  1. KRINKO/BfArM 2012, Bundesgesundheitsbl. 55:1244–1310 – Section 2.2.4.
  2. DIN EN ISO 11607-1/-2; DGSV guideline on validation of packaging processes.
  3. DIN 58953 (sterile supply management); routine operation checklist for sealing devices; MPBetreibV § 8.

How audit-ready is your practice?

Find out in just a few minutes how well prepared you are for the next inspection — category by category, with clear progress tracking and specific open items. Completely free with ClavioSoft.

ClavioSoft Audit-Ready Check: progress display and categories
Download ClavioSoft for free
Free forever For autoclaves of all manufacturers RKI-compliant

1 to 1 (from a total of 1)

Frequently Asked Questions

RKI Instrument Packaging – key questions answered clearly.

What is a sterile barrier system (SBS) in instrument reprocessing?
A sterile barrier system is the sealed packaging that protects the sterile instrument from recontamination until the point of use, such as a transparent sterilization pouch or a container. It must form a microbial barrier, allow steam penetration and permit aseptic opening. Requirements are governed by DIN EN ISO 11607-1.
Why must critical instruments be packaged before sterilization?
Packaging preserves the sterility of critical instruments until the point of use. It forms the sterile barrier system, which keeps out dust, moisture and micro-organisms after sterilization. Without packaging, sterility would no longer be guaranteed once the instruments are removed from the autoclave.
Must instruments be packaged or may I sterilize them unwrapped?
You may autoclave instruments wrapped, unwrapped or in a mixed load. Critical instruments that come into contact with blood or penetrate skin or mucous membranes must be packaged. Non-critical and semi-critical instruments may be sterilized unwrapped.
Which instruments are sterilized wrapped and which unwrapped?
Critical instruments are sterilized wrapped (sealed); semi-critical instruments may also be autoclaved unwrapped. The determining factor is the risk classification per KRINKO. In case of doubt, for example in a piercing studio, instruments are generally classified as critical and must be packaged.
Which standard must sterilization packaging meet?
Packaging material must meet DIN EN ISO 11607-1 or the DIN EN 868 series and must be declared suitable by the manufacturer for the chosen method (e.g. steam at 134 degrees C). The sealing process is validated in accordance with DIN EN ISO 11607-2.
Are sterilization pouches manufactured in accordance with DIN EN 868?
Yes, sterilization pouches are manufactured in accordance with the DIN EN 868 series. Together with DIN EN ISO 11607-1, this standard governs the requirements for materials and sterile barrier systems. Make sure the packaging is approved for steam sterilization.
What types of packaging are available for instrument sterilization?
Common types include sterilization paper rolls (with a sealing device), self-sealing pouches, sterilization containers and crepe or non-woven sheets. Rolls are suitable for larger sets, pouches for smaller instruments, containers for surgical sets and sheets for irregularly shaped instruments.
What are self-sealing pouches and do they require a sealing device?
Self-sealing pouches have an integrated adhesive strip and are closed by hand without a sealing device. They are suitable for occasional use and small quantities. For higher requirements, roll stock in combination with a sealing device is recommended.
When is a sealing device worthwhile instead of self-sealing pouches?
A sealing device with roll stock is worthwhile for higher instrument volumes, custom cut lengths and more stringent legal requirements. The seal seam is more consistent and the process can be validated in accordance with DIN EN ISO 11607-2. Self-sealing pouches are sufficient for occasional use.
How wide must the seal seam on sterilization pouches be?
The seal seam should be wrinkle-free and at least 6 mm wide. It must be continuous and firm, as it is the most critical point of the packaging. If the manufacturer specifies higher values, the manufacturer's specification applies.
What distance must be maintained between the seal seam and the instrument?
The distance between the seal seam and the instrument should be at least 30 mm (3 cm) so that no contact occurs during aseptic opening. If the manufacturer specifies a greater minimum distance, that specification applies. An overlap of at least 1 cm must also be maintained.
How long does packaged sterile goods remain sterile (sterile shelf life)?
The sterile shelf life depends on the packaging, storage conditions and manufacturer specifications; where applicable, it must be indicated on the packaging. For shelf lives exceeding six months, higher requirements apply to packaging quality and dust-free storage. The governing references are KRINKO/BfArM 2012 and DIN 58953.
How often must I perform a seal check on the sealing device?
During routine operation, inspect the seam daily using a seal check or ink test, visual inspection (uniform contrast, sharp contours) and peel test. Periodically, a tensile strength test of the seam is also performed. Document all inspections.
What is a peel test for the seal seam?
In a peel test, you open the packaging and check whether the paper separates from the film with virtually no residue. Fibre residue on the film indicates a correct seal. The peel test is part of the daily visual inspection of the seam.
Is a seal seam test (seal check) mandatory for sealing devices?
A seal seam test is required when you are sealing critical medical devices for use on patients. The exact interpretation may vary depending on the responsible authority, for example a daily seal check and annual validation with a seal seam tensile strength test. USB documentation is not mandatory everywhere.
Must sealing devices be validated?
Validatable sealing devices must undergo initial validation; in Germany, validation of the sealing process is indirectly mandatory via the relevant standards and the MPBetreibV. The governing standard is DIN EN ISO 11607-2. The exact interpretation depends on the responsible authority.
At what intervals must sealing devices be maintained?
In Germany, Austria and Switzerland, sealing devices are typically maintained every 24 months. Continuous sealing devices must additionally be maintained no later than after a specified number of sealing cycles. The manufacturer's specifications are authoritative.
Is a validatable sealing device required in podiatry?
Yes, in Germany a validatable sealing device is required in podiatry, even though this is not strictly enforced in every federal state. The reason is that podiatric instruments can penetrate the skin and are therefore classified as critical.
What is a sterilization container and what is it used for?
A sterilization container provides excellent protection for sets and surgical instruments. Instruments are wrapped in lint-free cloth, disposable filters are replaced before each use, and reusable filters are maintained and documented. Steam ports must remain clear; the container is sealed or tamper-proofed for storage.
How often must the filter of a sterilization container be replaced?
Disposable filters are replaced before each use. Reusable filters must be maintained and the maintenance documented. Only clean, intact filters ensure the microbial barrier and unobstructed steam access to the container.
What are crepe and non-woven sheets used for in packaging?
Crepe and non-woven sheets are flexible and ideal for irregularly shaped instruments and mixed sets. They are wrapped around the instruments. For sharp or pointed tips, protective caps should be used to prevent puncture.
What is a double sterile barrier system and when is it advisable?
A double sterile barrier system is a second outer wrap in addition to the inner SBS. It is advisable for particularly sensitive instruments or long-term storage. The additional protective packaging increases protection against recontamination.
How full may I fill sterilization pouches?
Do not overfill packages before sterilization so that steam can reach all surfaces. Overfilling can impede steam penetration and compromise the sterilization result. Leave sufficient space up to the seal seam.
May multiple instruments be placed in one sterilization pouch?
Yes, multiple instruments may be placed in one package provided the pouch is not overfilled and steam can reach all surfaces. Make sure the instruments do not damage each other. Secure heavy or pointed instruments with protective caps if necessary.
What is a process indicator or treatment indicator on the packaging?
A process indicator (treatment indicator) shows by colour change that the package has been through the sterilization process. It distinguishes processed from unprocessed sterile goods. Visual inspection of the indicator is part of the batch-related checks and the release procedure.
Must sterilization packaging have an indicator?
Use packaging with an integrated process indicator or apply one separately. It provides evidence that the package has been exposed to sterilization conditions. Class 1 process indicators are specified in the KRINKO/BfArM recommendation of 2012.
When do I label sterilization pouches, before or after autoclaving?
The labelling-relevant information is documented in the course of sterilization and release, not in advance during packing. During packaging, only non-critical preliminary labelling such as the contents is permitted, outside the seal seam and only on the paper side.
On which side of the sterilization pouch may I write?
Non-critical preliminary labelling during packing is applied outside the seal seam and only on the paper side. Never write over the seal seam or on the film in the instrument area, as this could compromise the microbial barrier. Labelling information follows after sterilization.
Which pen is suitable for labelling sterilization packaging?
Use a solvent-free sterilization marker without toxic chemicals. Write outside the seal seam on the paper side so that neither the seam nor the barrier is compromised. Such markers are recommended for packaging labelling.
What information must be included on the packaging label?
Labelling in the course of sterilization and release includes the sterilization date or batch identification, the expiry date or sterile shelf life where applicable, and the release decision. This information ensures traceability in accordance with KRINKO/BfArM 2012.
Must I re-inspect the packaging before each use?
Yes, regardless of batch-related checks, you must inspect the packaging again immediately before each use: dryness, integrity, intact seal seams and valid labelling. If the packaging is damaged, moist or the labelling is invalid, the instrument must not be used.
What do I do with a damaged or moist sterile package?
If the packaging is damaged, moist or the labelling is invalid, the instrument must not be used. It must be reprocessed and repackaged. A moistened package is no longer considered microbially sealed.
Must instruments be dry before packaging?
Yes, instruments should be thoroughly dried before packaging. Residual moisture can wet the packaging and destroy the microbial barrier. Dry instruments completely after cleaning and disinfection before sealing them.
What is a sterile goods packing workstation (SPA)?
SPA stands for Sterile Goods Packing Workstation (Sterilgut-Pack-Arbeitsplatz). This is where reprocessed instruments are inspected, assembled and packaged after cleaning and disinfection. The workstation is located in the clean area of the reprocessing unit.
Which programme do I select for wrapped instruments in the autoclave?
For wrapped instruments and hollow-body instruments, use the hollow programme (Universal 134 degrees C or 134 degrees C Hollow) of a Klasse B autoclave with fractionated vacuum. The solid programme for unwrapped instruments is not suitable for this purpose.
Can I sterilize wrapped instruments in a Class N autoclave?
No, Class N may only sterilize unwrapped, solid instruments. Wrapped, hollow or porous instruments require a Klasse B autoclave with a fractionated vacuum process. In human medicine, Klasse B is mandatory for critical instruments.
Must I package instruments in cosmetic foot care?
You must package instruments if they can puncture the skin or come into contact with blood, i.e. critical instruments. These are sealed and sterilized in a Klasse B autoclave. Non-critical instruments do not need to be packaged.
Which packaging is suitable for tattoo and piercing studios?
Self-sealing pouches are well suited for tattoo and piercing studios, as the instruments are generally classified as critical and packaged individually. Combined with a Klasse B autoclave, they ensure sterility. For higher volumes, roll stock with a sealing device is worthwhile.
How do I correctly package instrument sets for sterilization?
Package sets so that steam can reach all surfaces: do not place instruments too closely together, heavy items on the bottom, lighter ones on top. A photo of the standard loading and an instrument list are useful for documentation. Secure pointed instruments with protective caps to prevent puncture.
How do I protect pointed instruments from puncturing the packaging?
Use protective caps or protective paper to prevent puncture by sharp and pointed tips. Reusable protective caps must be approved by the manufacturer for multiple use and be autoclavable. Single-use caps may only be used once.
Are steam sterilization pouches also suitable for hot-air sterilizers?
No, self-sealing pouches for steam sterilization are not suitable for hot-air or dry-heat sterilizers. For critical instruments, a Klasse B steam autoclave is the governing standard in any case. Hot air requires packaging specifically approved for that method and is considered technically outdated.
What is the difference between a sterilization roll and a ready-made pouch?
Sterilization rolls (sold by the metre) can be individually cut to the instrument size and require a sealing device; they are suitable for larger instruments and sets. Ready-made self-sealing pouches are convenient for small quantities and are closed without a device.
Are there plain paper pouches without film for sterilization?
Paper pouches with a self-adhesive flap generally still have a transparent film side so the contents remain visible. For storage entirely without film, sterilization containers are suitable. For special material requirements, it is best to contact us directly.
How do I correctly store packaged sterile goods (paper or film side)?
Store packaged sterile goods upright, film to film and paper to paper, or with the paper side facing down. This keeps the microbial barrier protected and prevents damage to the packaging. Store in a dust-free, dry environment.
Must sterilization packaging allow steam to pass through?
Yes, the sterile barrier system must form a microbial barrier while allowing steam to pass through so that the instruments inside are sterilized. The paper side is steam-permeable, the film side maintains the microbial barrier. Both functions are governed by DIN EN ISO 11607-1.
How do I correctly load sterilization pouches into the autoclave?
Place pouches loosely with the paper side facing up; for vertical arrangement, film to film and paper to paper. This allows residual moisture to drain, the packaging dries better and steam and air penetrate the load evenly.
May expired packaged sterile goods be re-sterilized?
Packaged instruments may be re-sterilized after the shelf life has expired, provided the packaging is undamaged and clean, but they must be repackaged rather than sterilized in the same pouch. If the packaging is damaged or sterility can no longer be guaranteed, complete reprocessing is required.
How is the sealing process validated?
The sealing process is validated in accordance with DIN EN ISO 11607-2 by verifying and documenting the temperature, contact pressure and speed of the sealing device. Validation demonstrates that the device reproducibly creates tight seams. It complements the daily routine checks.
What should I look out for when packaging hollow-body instruments?
Hollow-body instruments are packaged individually and sterilized using the hollow programme of a Klasse B autoclave so that steam reaches the lumen. Steam penetration can be verified with a Helix test. Ensure the instruments are dry and the seal seam is intact.
How much does the initial validation of a validatable sealing device cost?
For a validatable sealing device, initial validation typically costs around 519 euros; subsequent revalidation is less expensive, at approximately 100 euros. Ideally, combine the maintenance and revalidation of your autoclave and sealing device in a single package.