The correct documentation for instrument preparation
How must instrument reprocessing be documented?
LEGAL BASIS OF INSTRUMENT REPROCESSING
The guidelines of the Robert Koch Institute (RKI) describe the legal basis for instrument reprocessing. This RKI guideline was created together with the Commission for Hospital Hygiene and Infection Prevention (KRINKO). Proper reprocessing is assumed according to the Medical Devices Operator Ordinance (§ 8 Para. 2) if the recommendation "Hygiene requirements for the reprocessing of medical devices" is observed. In the RKI recommendation, the process of instrument preparation is divided into the following 4 areas:- instrument classification
- treatment cycle
- documentation
- validation
- At least 5 years storage obligation according to RKI
- 5 or 10 years according to StGB §78
- 30 years according to BGB §199
In the case of instrument preparation, the burden of proof is reversed.
- A doctor (the operator) must always be able to prove that the instrument reprocessing was successful.
- The documentation is necessary to protect against claims for damages due to negligent action.
THE CORRECT DOCUMENTATION OF THE INSTRUMENT PREPARATION
The minimum documentation requirements for instrument reprocessing are described in device-specific standards.- EN 13060: The standard for small steam sterilizers (up to a chamber volume of 24 liters)
- EN 285: The standard for large steam sterilizers (from a chamber volume of 24 liters)
- EN 15883: The standard for thermal disinfectors
- EN 13060 stipulates that an automatic process assessment system must be integrated in the autoclave. This means that the autoclave continuously and automatically checks all relevant process parameters during the sterilization of instruments. It is therefore not necessary to manually check the process parameters such as the temperature reached and the holding time after sterilization. The autoclave must be able to do this itself.
- EN 13060 prescribes the presence of an error detection system. For this reason, several high-precision temperature sensors, at least one pressure sensor and a precise real-time clock are installed in autoclaves. The EN 13060 standard also describes which errors must be detected.
- EN 13060 stipulates that the autoclave must be able to create process documentation itself. Most autoclaves are equipped with a printer or USB port to be able to meet exactly this requirement of the standard. The documentation is thus printed out by the autoclave during the sterilization of instruments or saved on a stick.
Do the sterilization protocols to document the instrument preparation have to be printed out?
The RKI recommendation says very clearly: No. The RKI recommendation states that the sterilization protocols do not necessarily have to be printed out. The logs can also be saved on appropriate data carriers for backup purposes. The RKI (source: Robert Koch Institute "Hygiene requirements for the processing of medical products") writes: " The measured values of the process parameters recorded during the processing ... must be documented. The records may also be kept on image or data carriers ”.
The documentation options for Steri24 devices are:
Device | documentation options |
Class B Pro Autoclave |
|
Aqua-Plus thermal disinfector |
|
Data backups: The data on the USB stick should be backed up to a PC or hard drive at regular intervals (e.g. weekly). According to a survey by Steri24, 94% of clinics and practices in the DA-CH area use a USB stick to save the autoclave logs. Printing logs is not mandatory. In general, the complete documentation of the instrument preparation can be done completely paperless.
HOW DOES THE CORRECT BATCH CONTROL WORK FOR AUTOCLAVES IN INSTRUMENT PROCESSING?
If hollow bodies such as handpieces are also sterilized in the autoclave, then a helix test or PCD test (Process Challenge Device) is required. The aim of the test is to verify whether the sterilization of hollow bodies (e.g. handpieces) was successful. If no hollow bodies and only simply wrapped instruments are sterilized in the autoclave, a class 5 indicator is sufficient.How do I use the Helix test? The Helix test consists of a Helix test device into which a test strip can be inserted. The test device simulates the geometry of a hollow body. Before each sterilization process, a new test strip is inserted into the Helix test. The Helix test body is then placed in the autoclave together with the instruments and sterilized at the same time. After sterilization, a discoloration of the test strip indicates whether the sterilization was successful or not.
The manufacturer's recommendation of the class B Pro autoclaves from our shop for routine operation also correspond to the recommendations of the RKI. The recommendation is as follows:
Time schedule | Description of the exam |
---|---|
Every day the sterile is used | Visual inspection of the autoclave for damage Inspection of the autoclave display for warnings |
Once a week before starting work | Start of the "vacuum test" program |
With every sterilization cycle |
|
HOW DOES THE CORRECT BATCH RELEASE WORK WHEN PREPARING INSTRUMENTS?
The batch release phase begins after the autoclave has completed the sterilization cycle. "Approval for the use of instruments" must be carried out by trained personnel. We train your staff with a training certificate - contact us. Probably the most commonly used method of batch release is by means of a checklist and documentation in a steri book. You don't have to spend any money on it. We would be happy to send you a print template of the Steri book. You can print out the documents and file them in a folder. Contact us via contact form or chat and leave us your email address. The trained staff must carry out the following steps:
-
Process evaluation: The Steri24 are equipped with an automatic EN13060-compliant process evaluation. It is therefore sufficient to read the result of the sterilization protocol on the autoclave display. The process evaluation is carried out by the device itself without additional work for the personnel.
-
Documentation of batch indicators: The batch indicators (positive / negative) must be documented.
-
Visual check: A visual check must be carried out, for example for residual moisture or torn packaging. There is a separate checklist for this, which we will be happy to send you if required. Contact us.
-
Approval for renewed use: The release for renewed use must be carried out with a signature.
-
Labeling of packaged instruments: The packaged instruments must be labeled by specifying the autoclave, the current date, the use-by date, the batch number. and the person ID are designated. There is a standardized adhesive label gun for this.

Figure 1: Template for the documentation of the batch release
Labeling of the packaging after autoclaving is carried out using labeling tongs. The label must have at least the following content.
Personnel No. | Device No. | Batch number |
The "Use By" date is typically 6 months from the date of sterilization.